示波器差分探头 DP25
产品简介
. 3 RANGE ATTENUATOR
. 2 POWER SOURCE
. 1000V INPUT CAT III
. LOW BATT
. . AUTO OFF
. SEPARATING DESIGN
. CONVENIENT, DURABLE
. MEET CE, IEC 1010, EN 61010, UL 3111
产品详细信息
Band width (-3dB, 50Ω load): DC-25MHz . ( x 20: DC~15MHz)
Accuracy: ± 2% at20-30℃ 70% RH after 20 minutes warm up.
Attenuation: x 20, x50, x 200 ( Into 1MΩ scope ) ; x 40, x 10
, x 400 ( Into 50Ω load )
Maximum operation Voltage ( DC+ peak AC ) :
≦ ± 140V at x 20
≦ ± 350V at x 50
≦ ± 1300V at x 200
Maximum input DifferentialVoltage : 1300V( DC + peak AC ) or 1000V RMS
Maximum input Voltage toGround : 1000V ( DC + peak AC ) or 600V RMS
Common Mode Rejection Ratio (CMRR ):
60Hz: >10,000 : 1
100Hz: >1,000 :1
1MHz: > 300 :1
Noise:≦ 2 mV rms (Into 50Ω load )
Input Impedance:
2MΩ, 2.3 PF between inputs andground.
4MΩ, 1.2 PF between inputs.
Power:Oneinternal 9V alkaline battery or external 6V-9VDC.
WARNING
- Do not use DP-25 above 1000v ( DC + peak AC) between ground and the input or 1,300V ( DC + peak AC ) betweenthe input lead.
- Do not operate DP-25 in wetor damp condition.
- Do not operate DP-25 in anexplosive atmosphere.
- Do not immerse DP-25 in liquids.
- Do not operated DP-25 without covers.
- Please change the battery when the "LOW BATT " LED is lighted. At this time DP-25 can operate but not guaranteed the accuracy.
- DP-25 can not operate if bothPOWER and LOW BATT LED are not light.
FEATURES
- TheDP-25 FET input differential probe provieds a safe means ofmeasuring circuits with floating potentials up to 1000 V ( DC+ peakAC ) from ground and 1300V ( DC + peak AC )differential.
- TheDP-25 converts the high voltage differencial input signal to a lowvoltage ground refereced signal for display on anyOscilloscope.
- Theoutput BNC of DP-25 is calibrted to drive a high impedance (1MΩ ) load.
INSTRUCTION FORUSE
- Connect theoutput BNC of DP-25 to the input BNC of the Oscilloscope by theaccessory BNC cable.
- Adjustthe vertical offset of the Oscilloscope ifnecessary.
- Set theselect proper range of the DP-25 and the V / DIV of theOscilloscope according to the scale conversionchart.
- Scaleconversion chart : The effective V / DIV is the attenuation factorof x 20, x 50, x 200 multiplied by the scale factor of theOscilloscope. It will be twice when the 50 Ωload was used. For example, withthe range set at x 200,
and the scope set to 0.5 V / DIV, the effective V / DIV equals 200x 0.5 or 100 V, with the 50 Ω load wasused, it becomes 200V, the power consumption will increasetoo.
NOTE: If the voltageof the input signal exceeeds the linear range of the settingrange.The signal output of the DP-25 would not accurately, the waveform display will be cut off.