示波器差分探头 DP30HS
产品简介
. High Sensitivity Model
. 65 Vp-p/30MHz Max
. 3 Range: x1/ x2/ x10
. Input Impedance (Between Input): 2MΩ // 1.7PF
产品详细信息
DP-30HS HIGH SENSITYVITY DIFFERENTIAL PROBE SPECIFICATIONS
(1) Bandwidth:
DC - to 30 MHz (-3 dB) for x 2,x10
DC - to 15 MHz (for attenuation x 1)
(2) Attenuation: x 1, x 2, or x 10
(3) Accuracy: +/- 2%
(4) Voltage Input Ranges (DC + AC peak to peak)
≦ +/- 6.5 V for x 1, (i.e about 2.3VRMS or DC)
≦ +/- 13 V for x 2, (i.e about 4.6VRMS or DC)
≦ +/- 65 V for x 10, (i.e about 23 V RMSorDC)
(5) Permitted Max Input Voltage
Max differential voltage: 65 V (DC +ACpeak to peak)
Max voltage between each input terminalandground: 600 V RMS
(6) Input Impedance:
Differential: 2 MΩ / 1.7 pF
Between terminals and ground: 1 MΩ /3.4pF
(7) Output: ≦ +/- 7.0 V
(8) Output Impedance: 50 Ω
(9) Rise Time: 12 ns for x 2, and x 10; 24 ns for x 1
(10) Rejection Rate on Common Mode:
60 Hz: > 80 dB ; 100 Hz: > 60 dB ; 1 MHz:>50 dB
(11) EXT. Power Supply: External 9 V DC power supply.
(12) Consumption: 35 mA max (0.4 WATT)
(1) Bandwidth:
DC - to 30 MHz (-3 dB) for x 2,x10
DC - to 15 MHz (for attenuation x 1)
(2) Attenuation: x 1, x 2, or x 10
(3) Accuracy: +/- 2%
(4) Voltage Input Ranges (DC + AC peak to peak)
≦ +/- 6.5 V for x 1, (i.e about 2.3VRMS or DC)
≦ +/- 13 V for x 2, (i.e about 4.6VRMS or DC)
≦ +/- 65 V for x 10, (i.e about 23 V RMSorDC)
(5) Permitted Max Input Voltage
Max differential voltage: 65 V (DC +ACpeak to peak)
Max voltage between each input terminalandground: 600 V RMS
(6) Input Impedance:
Differential: 2 MΩ / 1.7 pF
Between terminals and ground: 1 MΩ /3.4pF
(7) Output: ≦ +/- 7.0 V
(8) Output Impedance: 50 Ω
(9) Rise Time: 12 ns for x 2, and x 10; 24 ns for x 1
(10) Rejection Rate on Common Mode:
60 Hz: > 80 dB ; 100 Hz: > 60 dB ; 1 MHz:>50 dB
(11) EXT. Power Supply: External 9 V DC power supply.
(12) Consumption: 35 mA max (0.4 WATT)