TH2828/TH2828A Precision LCR Meter TH2828/TH2828A
产品简介
TH2828 series component measurement meter is 1 new-generation impedance measurement instruments applying the advanced technique of auto balance bridge in the world .It can be used for all component an
产品详细信息
TH2828/TH2828A Precision LCR Meter Specifications
Measurement function | |||||
Measurement parameters | 22 parameter combinations, including |Z|, |Y|, C, L, X, B, R, G, D, Q, θ, ESR (equivalent series resistance), Rp (equivalent parallel resistance) | ||||
Equivalent circuit | Series, Parallel | ||||
Mathematical function | ΔABS, Δ% | ||||
Mode | Auto, Hold, Manual | ||||
Subsection | 9 sects: 10Ω, 30Ω, 100Ω, 300Ω, 1kΩ, 3kΩ, 10kΩ, 30kΩ, 100kΩ | ||||
Trigger mode | Internal, Manual, External, BUS | ||||
Measurement Time (≥1kHz) | Fast: 32ms(25ms@1MHz), Medium: 90ms, Slow: 650ms | ||||
Average time | 1—255 | ||||
Delay time | 0—60s, with step of 1ms | ||||
Calibration function | Open/Short frequency pint, full frequency correction, Load correction | ||||
Measurement terminal | 4-terminal pair | ||||
Test cable length | Standard: 0m, 1m Option: 2m, 4m | ||||
Display mode | Direct, Δ, Δ%, bin No, bin counter, list sweep, V/I (voltage/current monitor) | ||||
Display | 320×240 dot-matrix graphic LCD display | ||||
Measurement signal | |||||
TH2828 | More than 6,000 points from 20 Hz to 1MHz | ||||
TH2828A | 44 points from 50Hz to 1MHz: 50,60,80,100,120,150,200,250,300,400,500,600,800,1k,1.2k,1.5k,2k2.5k,3k,4k,5k,6k,8k,10k,12k,15k,20k,25k ,30k,40k,50k,60k,80k,100k,120k,150k,200k,250k,300k,400k,500k,600k,800k,1M | ||||
TH2828S | 20 Hz-1MHz, with min. resolution of 1mHz, randomly programming | ||||
accuracy | 0.01% | ||||
Output impedance | 30 Ω , 100 Ω selectable only 100 Ω available when using TH10301 | ||||
Normal | When terminal is open or short, signal current or voltage can be programmed to se | ||||
Constant level | Ensure that voltage or current of DUT doesn’t change with impedance of DUT | ||||
Normal V Normal I | 5mVrms — 2Vrms 50μArms — 20mArms | ||||
Constant level V Constant level I | 10mVrms — 1Vrms 100μArms — 10mArms | ||||
Normal V Normal I | 5mVrms — 20Vrms 50μArms — 200mArms | ||||
Constant level V Constant level I | 10mVrms — 10Vrms 100μArms—100mArms | ||||
Standard | 0V, 1.5V, 2V DC | ||||
Range | Resolution | ||||
±(0.000 — 4.000)V DC | 1mV | ||||
±(4.002 — 8.000)V DC | 2mV | ||||
±(8.005 — 20.000)V DC | 5mV | ||||
Measurement display range | |||||
|Z|, R, X | 0. 01mΩ — 99.9999MΩ | ||||
|Y|, G, B | 0. 01nS — 99.9999S | ||||
C | 0. 01pF — 9.99999F | ||||
L | 0.01nH — 99.9999kH | ||||
D | 0.00001 — 9.99999 | ||||
Q | 0.00001 — 99999.9 | ||||
Θ (angel DEG | -179.999º — 179.999 º | ||||
θ (radian RAD) | -3.14159 — 3.14159 | ||||
Δ% | -999.999% — 999.999% |
List sweep | |||
Max. 10 measurement signal frequencies and levels can be selected to scan. Scanning can be performed at single or continues mode. When using TH10301, list sweep of DC bias level can be performed. | |||
Comparator and interface | |||
Comparator | 10-bin sorting and bin counter for measurement parameters IN/OUT judgment for sub parameters | ||
Bin counter | 0—999999 | ||
List sweep comparator | HIGH/IN/OUT judgment for any point can be performed. | ||
Input protection | |||
Internal circuit protection, when a charged capacitor is connected to the Unknown terminals. The maximum capacitor voltage is: C in Farads | |||
Others | |||
Memory | 20 sets of instrument setting files saved and recalled in non- volatile memory. Another 40 sets of settings saved in USB disk (not for TH2818) | ||
GPIB,RS232C | All instrument control settings, measured values, comparator limits and list sweep tables can communicating with computer or other instruments by GPIB (optional for TH2828A ) or RS232C. | ||
Options | |||
TH10301 | Signal power amplifier/DC bias. Increasing AC measurement signal to 20Vrms/100mArms. Extending DC bias range to 40V/100mA | ||
TH10401 | Extending test cable to 2m/4m | ||
TH10202 | Handler interface Nine sets of High/Low limits can be input, allowing 10-bin sorting for L, C, or │Z│. The handler interface enables systemization with an automatic component sorting machine. All signals are optically isolated. | ||
Accuracy(Refer to manual for more details) | |||
Measurement conditions | Warm-up time | ≥30 minutes | |
Surroundings temperature | 23±5ºC | ||
Signal voltage | 0.3Vrms – 1Vrms | ||
Calibration | Open, Short | ||
Test cable length | 0m | ||
|Z|, |Y|, C, L, X, B, R, G, | Ae = ±[A+(Ka+Kb+Kc)×100] (% of reading) 1. A is basic accuracy factor as in figure 1 and 2 2. Ka, Kb are impedance proportional factors 3. Kc is calibration interpolation. Direct correction frequencies: Kc=0, Other frequencies Kc=0.0003 4. D ≤ 0.1, for C, L, B measurement Q ≤ 0.1, for R, G measurement | ||
D | ±[Ae/100] (direct reading of D) Here, A=[A+(Ka+Kb+Kc)×100] | ||
Q When (Qx×De<0.1) | Here, Qx is measured Q value, De is D’s accuracy | ||
θ | DEG | ±[Ae/100] (direct radian) | |
RAD | ±[(180/π)×(Ae/100)] (direct angle) |
General Specifications
Operating temperature & humidity | 0°C-40°C, £90%RH | |
Voltage | 99V-121V,198V-242V | |
Frequency | 47.5Hz-63Hz | |
Power consumption | ≤ 100VA | |
Dimensions (W×H×D) | 430mm×185mm×473mm | |
Weight | Approx. 15 kg |
Ordering Information
TH2828 Precision LCR Meter
TH2828A Wide-frequency LCR Meter
TH2828S Automatic Component Analyzer
Instrument Accessories
TH26005-1 4-terminal test fixture
TH26011A 4-terminal pair Kelvin test cable
TH26010 gilded shorting plate
TH10002 GPIB interface board (only for TH2828/TH2828S)
TH26025 USB interface adapter
TH26026 USB memory card
Options
TH26005 4-terminal test fixture
TH26005-2 4-terminal test fixture
TH26006 axial component test fixture (used with TH26005)
TH26007 core test fixture
TH26008 SMD component test fixture
TH26009 SMD component test tweezers
TH26029A SMD component test tweezers
TH10301 20Vrms/40V DC power amplifier/DC bias board
TH10401 2m/4mcable length operation (only for TH2828)
TH10002 GPIB interface board
TH10202 Handler/Scanner interface board
TH11014 TH2828/A GPIB control software
TH11015 TH2828S GPIB control software
TH12019 TH2828/A RS232C control software
TH12020 TH2828S RS232C control software