TH2825 High-speed LCR Meter TH2825
产品简介
TH2825 High-speed LCR Meter is new-generation component parameter measurement instrument. It offers high measurement speed up to 15times/second, 10 frequencies from 40Hz to 100kHz
产品详细信息
TH2825 High-speed LCR Meter Specifications
| TH2825 | TH2825A | |
Measurement function | |||
Parameter | LCR | |Z|,R, X, G, B, C, L, Q, D, θ(deg), θ(rad) | |
Transformer | ------ | DCR, DCR2,M, Turns, N,1/N,L2A,L2B | |
Basic accuracy | 0.1% | ||
Equivalent circuit | Series, parallel | ||
Mathematical function | ΔABS (absolute value deviation), Δ% (percent deviation) | ||
Range mode | Auto, Hold, Manual conversion | ||
Trigger mode | Internal, Manual, External, BUS | ||
Measurement time | Fast: 15ms (note 1), medium: 66ms, slow: 500ms | ||
Average times | 1—255 | ||
Calibration function | Open, short, load correction | ||
Measurement terminal | 5-terminal | ||
List sweep | 4 frequencies, level, DC bias list sweep measurement | ||
Display mode | Direct, ΔABS, Δ%, V/I (voltage/current monitor), bin No., and bin counter | ||
Display | 240×64 dot-matrix LCD display with resolution of 5 digits | ||
Measurement signal | |||
Signal frequency | Total 10 points: 50Hz,60Hz,100Hz,120Hz,1kHz,10kHz,20kHz, 40kHz, 50kHz,100kHz | ||
Output impedance | 25Ω,50Ω,100Ω, 25Ω/100Ω,C.V. (constant voltage) | ||
Test level | 10mVrms—1.0Vrms, with step of 1mV | ||
DC bias source | ------ | Fixed ±2VDC,0~±5VDC adjustable | |
Resistance is 100Ω: ±50mA Max Resistance is 25Ω: ±200mA Max | |||
Measurement display range | |||
|Z|, R, X | 0.0001Ω — 99.999MΩ | ||
G, B | 0.0001μS — 999.99S | ||
C | 0.001pF — 1.9999F | ||
L | 0.001µH — 99.99kH | ||
D | 0.0001 — 9.9999 | ||
Q | 0.0001 — 99999 | ||
θ (DEG) | -179.99º — 179.99 º | ||
θ (RAD) | -3.1416 — 3.1416 | ||
Δ% | -99.999% — 999.999% | ||
N,1/N | ------ | 0.001—999.99 | |
DCR | ------ | 0.1 mΩ—199.99 kΩ | |
M,L2 | ------ | 0.001 μH—99.99 kH | |
Comparator and interface | |||
Comparator | 10 bins: 8 pass bins, 1 fail bin, 1 AUX bin | ||
HIGH/PASS/LOW | |||
Memory | 12 sets of instrument settings saved and loaded | ||
Interface | RS-232C, HANDLER, GPIB (option) |
Note 1: Fast measurement time includes A/D, calculation, main/sub parameter comparison judgment,and
small-character display of measurement parameter. If measurement parameter is displayed in large
character, fastest measurement time is added by 5ms.
General Specifications
Operating temperature | 0°C-40°C | |
Operating humidity | £90%RH | |
Voltage | 99V-242V | |
Frequency | 47.5Hz-63Hz | |
Power consumption | ≤ 30VA | |
Dimensions (W×H×D) | 350mm×122mm×418mm | |
Weight | Approx. 4.5 kg |
Ordering Information
TH2825 High-speed LCR Meter
TH2825A High-speed LCR Meter
Instrument Accessories
TH26005 4-terminal test fixture
TH26011 4-terminal test fixture 4-terminal Kelvin test cable
TH26010 gilded shorting plate
Options
TH26001 4-terminal test fixture
TH26005-2 4-terminal test fixture
TH26006 axial component test fixture (used with TH26005)
TH26007 core test fixture
TH26008 SMD component test fixture
TH26009 SMD component test tweezers
TH26029A SMD component test tweezers
TH26012 transformer test fixture (only for TH2825A)
TH26004-1 4-terminal test fixture
TH10001 GPIB interface board